Suunto DX has the following dive modes:
By default, Air mode is activated when you enter dive mode. You can change which mode is activated or turn dive mode off under the general settings.
To change dive modes:
Each dive mode has its own settings which you need to adjust while in the given mode.
To modify dive mode settings:
Some settings cannot be changed until five (5) minutes have elapsed after the dive.
Air mode is for diving with regular air and has the following settings:
Suunto DX has a mixed dive mode for diving with oxygen and/or helium mixture in an open system.
Mixed mode has the following settings available:
In mixed mode, the default setting is standard air (21% O2 and 0% He), and oxygen partial pressure (PO2) is 1.4 bar (20 psi).
The default setting for maximum oxygen partial pressure is 1.4 bar (20 psi). Valid values are 0.5-1.6 bar (7-23 psi).
If you have using more than one gas on a dive, Suunto DX allows you to change enabled gas mixes during the dive.
A dive is always started with Mix1. You can change to another enabled mix that is within the set maximum oxygen partial pressure. The tissue calculation during the dive is based on the mixes you have selected as Primary gases.
To change gases during a dive:
If no button is pressed within 15 seconds, the dive computer goes back to the dive display without changing the gas mix.
The mix number, O2% and PO2 for the mixes are shown when scrolling. If the set PO2 limit is exceeded, the PO2 value blinks. You cannot change to the gas in this case. The mix is shown, but you cannot select it to use.
When ascending, Suunto DX prompts you to change gas when the PO2 level you have set for the next mix allows a gas change.
CCR mode is a dive mode dedicated to rebreather diving.
The CCR mode has the following settings:
In CCR mode you can define up to three diluent gases and up to eight bailout gas mixes. Use the setpoint settings to define your high and low set points and switch depths (see Setpoints).
On a rebreather dive, you need at minimum two closed-circuit gases: one is your pure oxygen tank, and the other is a diluent.
The correct oxygen and helium percentages of the diluent gas(es) in your diluent cylinder(s) must always be entered into the dive computer (or through DM5) to ensure correct tissue and oxygen calculation. Diluent gas(es) used on a rebreather dive are found under CC gases in the main menu.
To modify diluent gases
As with closed-circuit gases, you must always define the correct oxygen and helium percentages of open-circuit gases (bailout gases) to ensure correct tissue and oxygen calculation.
Follow the same procedure as the CC gases to define your open-circuit gases under the OC Gases settings.
After entering values for Mix1, you can create additional mixes,Mix2 to Mix8. Each additional mix can be primary, secondary, or off. Mix1 is always set as a primary gas.
To minimize the risk of error during a dive, it is highly recommended that the mixes are set in the proper order. This means that as the mix number rises, so does oxygen content. This is the order they are usually used during the dive. Before a dive, only enable the mixes you actually have available and remember to check the set values to ensure they are correct.
The ascent time is calculated based on the assumption that you start the ascent profile immediately and all the primary gases are changed as soon as their maximum operating depth allows it. That is, using the gases that are set as primary, the most optimal ascent schedule for the moment is calculated.
To see the most pessimistic ascent schedule, that is, a schedule for the situation when gases are not changed at all, you can set the gases as secondary and the time it takes to finish the decompression using the current breathing gas is shown as the ascent time.
Showing the most pessimistic ascent schedule during a long dive may result in the ascent time no longer fitting the reserved field, and the dive watch displays '––'.
While setting the gases, notice that the calculated maximum operating depth is displayed in the upper field. You cannot change to this gas before you have ascended above this depth.
In CCR mode there are two setpoint values that you can set, low and high. Typically you do not need to modify the default setpoint values. However, you can change them as needed either in DM5 or under the CCR mode settings.
To change set point values:
You can use automatic or manual setpoint switching. The automatic low setpoint switch depth is 4.5 m (15 ft) by default, and the high setpoint switch depth is 21 m (70 ft).
The auto setpoint switching is off by default for the low setpoint and on for the high setpoint.
To modify setpoint switching:
The high and low setpoints or a custom (manual) set point can be adjusted during a dive.
To modify setpoints during a dive:
In CCR dive mode, Suunto DX allows both setpoint and gas changes to enabled gas mixes during the dive.
To adjust the diluent during a dive:
Enabled open-circuit gas changes are made in the same way and at the same time as the closed-circuit (CC) dive changes to an open-circuit (OC) dive. This feature is useful in bail-out situations.
Suunto DX displays the text HYPER if the diluent gas PO2 at the current depth is over 1.6. The text HYPOX is displayed if the diluent gas PO2 at the current depth is less than 0.18.
With Gauge mode, you can use Suunto DX as a bottom timer.
The timer in the center of the display shows dive time in minutes and seconds and activates at the start of the dive. The total running dive time, in minutes, is in the lower right corner.
The timer in the center of the display can be used as a stopwatch by pressing
during the dive.Pressing
resets the main timer and adds a bookmark to the dive log. The previously-timed interval is displayed below the main timer.Gauge mode has the following settings
Gauge mode is a bottom timer only and thus includes no decompression information or calculations.